Animal-Based Model to Investigate the Minimum Tumor Size Detectable with an Electrical Impedance Scanning Technique

  • Ansgar Malich, MD

      Affiliations

    • Institute of Diagnostic and Interventional Radiology, Friedrich-Schiller University Jena, Bachstrasse 18, 07740 Jena, Germany
    • Corresponding Author InformationInstitute of Diagnostic and Interventional Radiology, Friedrich-Schiller University Jena, Bachstrasse 18, 07740 Jena, Germany
  • ,
  • Thomas Böhm, MD

      Affiliations

    • Institute of Diagnostic Radiology, University of Zurich, Switzerland
  • ,
  • Tobias Fritsch, MD

      Affiliations

    • Institute of Diagnostic and Interventional Radiology, Friedrich-Schiller University Jena, Bachstrasse 18, 07740 Jena, Germany
  • ,
  • Mirjam Facius, MD

      Affiliations

    • Institute of Diagnostic and Interventional Radiology, Friedrich-Schiller University Jena, Bachstrasse 18, 07740 Jena, Germany
  • ,
  • Martin G. Freesmeyer, MD

      Affiliations

    • Institute of Diagnostic and Interventional Radiology, Friedrich-Schiller University Jena, Bachstrasse 18, 07740 Jena, Germany
  • ,
  • Roselle Anderson

      Affiliations

    • Department of Special Systems-Women's Health, Siemens-Elema AB, Solna, Sweden
  • ,
  • Marlies Fleck, MD

      Affiliations

    • Institute of Diagnostic and Interventional Radiology, Friedrich-Schiller University Jena, Bachstrasse 18, 07740 Jena, Germany
  • ,
  • Werner A. Kaiser, MD

      Affiliations

    • Institute of Diagnostic and Interventional Radiology, Friedrich-Schiller University Jena, Bachstrasse 18, 07740 Jena, Germany

Received 15 July 2002; received in revised form 23 August 2002; accepted 26 August 2002.

Abstract 

Rationale and Objectives

The purpose of this study was to determine the minimum tumor size detectable with electrical impedance scanning (EIS) in laboratory animals.

Materials and Methods

VX2 tumor cells (1 mm3) were implanted bilaterally into the upper leg musculature of five white New Zealand rabbits. EIS and ultrasound (US) were performed before, during, and immediately after implantation and on every 2nd day thereafter until tumors could be visualized with both modalities. This was followed by an extended follow-up regimen until a tumor size of 1 cm3 was reached. Rabbits were anesthetized subcutaneously.

Results

Tumors could not be implanted in one rabbit. Neither EIS nor US were performed in one rabbit due to severe skin alterations at the implantation site. No focal increase in conductance was visible before or immediately after tumor injection. The smallest tumor identified with EIS was determined with US to be approximately 8 mm3. The mean tumor size at initial detection was 52 mm3 (determined with EIS). In two cases, the tumor was first seen with US (EIS detection was delayed with a delay of 1 and 2 days, respectively). In all remaining cases, tumors were detected simultaneously with both EIS and US. All tumors were visible with EIS.

Conclusion

The animal-based model is feasible. VX2 tumors are detectable with EIS. Tumors characterized by a focal spot at EIS could be detected starting at a tumor size of 7.5 mm3.

Keywords: Breast neoplasms

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1076-6332(03)80786-9

doi:10.1016/S1076-6332(03)80786-9

Academic Radiology
Volume 10, Issue 1 , Pages 37-44, January 2003