Academic Radiology
Volume 15, Issue 3 , Pages 370-382, March 2008

Image Browsing in Slow Medical Liquid Crystal Displays1

CDRH/NIBIB Laboratory for the Assessment of Medical Imaging Systems, Division of Imaging and Applied Mathematics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, 10903 New Hampshire Avenue, Building 62, Room 3113, Silver Spring, MD 20993.

Received 9 September 2007; accepted 20 October 2007.

Rationale and Objectives

Statistics show that radiologists are reading more studies than ever before, creating the challenge of interpreting an increasing number of images without compromising diagnostic performance. Stack-mode image display has the potential to allow radiologists to browse large three-dimensional (3D) datasets at refresh rates as high as 30 images/second. In this framework, the slow temporal response of liquid crystal displays (LCDs) can compromise the image quality when the images are browsed in a fast sequence.

Materials and Methods

In this article, we report on the effect of the LCD response time at different image browsing speeds based on the performance of a contrast-sensitive channelized-hoteling observer. A stack of simulated 3D clustered lumpy background images with a designer nodule to be detected is used. The effect of different browsing speeds is calculated with LCD temporal response measurements from our previous work. The image set is then analyzed by the model observer, which has been shown to predict human detection performance in Gaussian and non-Gaussian lumpy backgrounds. This methodology allows us to quantify the effect of slow temporal response of medical liquid crystal displays on the performance of the anthropomorphic observers.

Results

We find that the slow temporal response of the display device greatly affects lesion contrast and observer performance. A detectability decrease of more than 40% could be caused by the slow response of the display.

Conclusions

After validation with human observers, this methodology can be applied to more realistic background data with the goal of providing recommendations for the browsing speed of large volumetric image datasets (from computed tomography, magnetic resonance, or tomosynthesis) when read in stack-mode.

Key Words: Model observer, display quality, contrast sensitivity, temporal response

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

1 The mention of commercial products herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services. This is a contribution of the US Food and Drug Administration, and is not subject to copyright.

PII: S1076-6332(07)00627-7

doi:10.1016/j.acra.2007.10.017

Academic Radiology
Volume 15, Issue 3 , Pages 370-382, March 2008