Academic Radiology
Volume 16, Issue 10 , Pages 1241-1250, October 2009

Is Coronary Stent Assessment Improved with Spectral Analysis of Dual Energy CT?

  • Ethan J. Halpern, MD

      Affiliations

    • Department of Radiology, Thomas Jefferson University, 132 S 10th Street, Philadelphia, PA 19107-5244
    • Corresponding Author InformationAddress correspondence to: E.J.H.
  • ,
  • David J. Halpern

      Affiliations

    • Department of Radiology, Thomas Jefferson University, 132 S 10th Street, Philadelphia, PA 19107-5244
  • ,
  • Jeffrey H. Yanof, PhD

      Affiliations

    • Philips Medical Systems, Cleveland, OH
  • ,
  • Sigal Amin-Spector, PhD

      Affiliations

    • Philips Medical Systems, Cleveland, OH
  • ,
  • David Fischman, MD

      Affiliations

    • Division of Cardiology, Department of Internal Medicine, Thomas Jefferson University, 132 S 10th Street, Philadelphia, PA 19107-5244
  • ,
  • Galit Aviram, MD

      Affiliations

    • Department of Radiology, Tel Aviv Sourasky Medical Center, Tel Aviv, Israel
  • ,
  • Jacob Sosna, MD

      Affiliations

    • Department of Radiology, Hadassah Hebrew University Hospital, Jerusalem, Israel

Received 24 March 2009; accepted 21 April 2009. published online 15 June 2009.

Rationale and Objectives

The aims of this study were to distinguish stents from iodinated contrast on the basis of spectral characteristics on dual-energy computed tomographic (DECT) imaging and to determine whether DECT imaging might provide a more accurate measurement of true stent lumen.

Materials and Methods

Three stainless steel stents and one cobalt chromium stent were scanned using a multidetector, single-source DECT scanner. Stents 2.5, 3.5, and 4.0 mm in diameter were filled with iodinated contrast, submerged in water, and scanned. Spectral analysis was performed to assess the separation of stents from iodinated contrast. Two independent reviewers measured stent lumen diameter and strut thickness on low-energy (L0), high-energy (L1), and combined-energy (Lc) images. Dual-energy full-width half-maximum edge detection analysis was used to provide an independent assessment of stent luminal diameter and strut thickness.

Results

Two-dimensional graphical plots of computed tomographic attenuation for the L0 and L1 images did not demonstrate a sharp separation between the absorption characteristics of stents and iodinated contrast material. Stent lumens were underestimated by approximately 50% on Lc images. Observer measurements on L1 images demonstrated a 24% decrease in strut thickness and a 25% increase in stent luminal diameter compared to L0 images (P < .0001). Full-width half-maximum measurements did not demonstrate significant changes in stent luminal diameters or strut thicknesses between L0 and L1 images.

Conclusions

Spectral analysis did not clearly distinguish stents from iodinated contrast with the DECT system used in this study. The larger stent lumens visualized by the high-energy components of the x-ray spectrum were not related to improved computed tomographic delineation of stent thickness.

Key Words: Coronary artery, coronary stent, CT angiography, dual energy CT, spectral classification

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1076-6332(09)00266-9

doi:10.1016/j.acra.2009.04.011

Academic Radiology
Volume 16, Issue 10 , Pages 1241-1250, October 2009