Academic Radiology
Volume 17, Issue 10 , Pages 1259-1266, October 2010

Adaptive Statistical Iterative Reconstruction Technique for Pulmonary CT:

Image Quality of the Cadaveric Lung on Standard- and Reduced-Dose CT

  • Masahiro Yanagawa, MD, PhD

      Affiliations

    • Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka, 565-0871, Japan
    • Corresponding Author InformationAddress correspondence to: M.Y.
  • ,
  • Osamu Honda, MD, PhD

      Affiliations

    • Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka, 565-0871, Japan
  • ,
  • Shigeyuki Yoshida, MD

      Affiliations

    • Department of Radiology, Minoh City Hospital, Minoh-city, Osaka, Japan
  • ,
  • Ayano Kikuyama, MD

      Affiliations

    • Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka, 565-0871, Japan
  • ,
  • Atsuo Inoue, MD, PhD

      Affiliations

    • Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka, 565-0871, Japan
  • ,
  • Hiromitsu Sumikawa, MD, PhD

      Affiliations

    • Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka, 565-0871, Japan
  • ,
  • Mitsuhiro Koyama, MD, PhD

      Affiliations

    • Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka, 565-0871, Japan
  • ,
  • Noriyuki Tomiyama, MD, PhD

      Affiliations

    • Department of Radiology, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita-city, Osaka, 565-0871, Japan

Received 31 March 2010; accepted 19 May 2010. published online 16 July 2010.

Rationale and Objectives

To evaluate thin-section computed tomography (CT) images of the lung reconstructed using adaptive statistical iterative reconstruction (ASIR) on standard- and reduced-dose CT.

Materials and Methods

Eleven cadaveric lungs were scanned by multidetector-row CT with two different tube currents (standard dose, 400 mA; reduced dose, 10 mA). The degree of ASIR was classified into six different levels: 0% (non-ASIR), 20%, 40%, 60%, 80%, and 100% (maximum-ASIR). The ASIR (20%, 60%, and 100%) images were compared with the ASIR (0%) images and assessed visually by three independent observers for image quality using a 7-point scale. The evaluation items included abnormal CT findings, normal lung structures, and subjective visual noise. The median scores assigned by the three observers were analyzed statistically. Quantitative noise was calculated by measuring the standard deviation in a circular region of interest on each selected image of ASIR (0%–100%).

Results

On standard-dose CT, the overall image quality significantly improved with increasing degree of ASIR (P ≤ .009, Wilcoxon signed-ranks test with Bonferroni correction). As ASIR increased, however, intralobular reticular opacities and peripheral vessels tended to be obscure. On reduced-dose CT, the overall image quality of ASIR (100%) was significantly better than that of ASIR (20%) (P ≤ .009). As ASIR increased, however, intralobular reticular opacities tended to be obscure. Using ASIR significantly reduced subjective and quantitative image noise on both standard- and reduced-dose CT (P < .001, Bonferroni/Dunn's method).

Conclusion

ASIR improves the image quality by decreasing image noise. Maximum-ASIR may be needed for improving image quality on highly reduced-dose CT. However, excessive ASIR may obscure subtle shadows.

Key Words: Computed tomography (CT), adaptive statistical iterative reconstruction (ASIR), cadaveric lung

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S1076-6332(10)00293-X

doi:10.1016/j.acra.2010.05.014

Academic Radiology
Volume 17, Issue 10 , Pages 1259-1266, October 2010