Liver Iodine Quantification With Photon-Counting Detector CT: Accuracy in an Abdominal Phantom and Feasibility in Patients

      Rationale and Objectives

      To compare the accuracy of iodine quantification in liver parenchyma and lesions between dual-source photon-counting detector CT (PCD-CT) and dual-source energy-integrating detector CT (EID-CT) in a phantom and to demonstrate the feasibility of iodine quantification with PCD-CT in liver parenchyma and lesions in patients.

      Materials and Methods

      An anthropomorphic abdominal phantom with a liver insert containing parenchyma and lesions was imaged on a clinical PCD-CT at 120kV and in the dual-energy mode on an EID-CT with kV-combinations of 80/Sn150kV, 90/Sn150kV, and 100/Sn150kV. Three patient sizes were imaged at three different radiation doses (CTDIvol: 5, 10, 15mGy). Thirty patients with liver cysts, hemangiomas or metastases imaged with PCD-CT were retrospectively included. Iodine maps were reconstructed and iodine concentrations were measured in liver parenchyma and lesions. For the phantom, iodine error was quantified as the absolute difference to the vendor's specifications as reference.

      Results

      Overall iodine error was 0.33 ± 0.29, 0.34 ± 0.32, 0.39 ± 0.37, 0.35 ± 0.39 mgI/mL for 80/Sn150kV, 90/Sn150kV, 100/Sn150kV of EID-CT, and PCD-CT, respectively, without significant differences between PCD-CT and EID-CT (p > 0.05). Radiation dose did not significantly influence error of PCD-CT (p > 0.05) nor EID-CT (p > 0.05). For both scanners, smaller patient sizes were associated with lower errors (p < 0.05). Iodine concentration and base material attenuation significantly influenced quantification for EID-CT (p < 0.05) but not PCD-CT (p > 0.05). In patients, iodine quantification was feasible in liver parenchyma, cysts, hemangiomas, and metastases.

      Conclusion

      Iodine quantification with PCD-CT is accurate in simulated liver parenchyma and lesions irrespective of radiation dose, iodine concentration, and base attenuation and is feasible in common liver lesions in patients.

      Key Words

      Abbreviations:

      EID-CT (energy-integrating detector CT), DECT (dual-energy computed tomography), PCD-CT (photon-counting detector CT)
      To read this article in full you will need to make a payment

      Purchase one-time access:

      Academic & Personal: 24 hour online accessCorporate R&D Professionals: 24 hour online access
      One-time access price info
      • For academic or personal research use, select 'Academic and Personal'
      • For corporate R&D use, select 'Corporate R&D Professionals'

      Subscribe:

      Subscribe to Academic Radiology
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Huda F
        • LeBedis CA
        • Qureshi MM
        • et al.
        Acute cholecystitis: diagnostic value of dual-energy CT-derived iodine map and low-keV virtual monoenergetic images.
        Abdom Radiol. 2021; 46: 5125-5133https://doi.org/10.1007/s00261-021-03202-9
        • Obmann MM
        • Punjabi G
        • Obmann VC
        • et al.
        Dual-energy CT of acute bowel ischemia.
        Abdom Radiol. 2021; https://doi.org/10.1007/s00261-021-03188-4
        • Cao J
        • Lennartz S
        • Parakh A
        • et al.
        Dual-layer dual-energy CT for characterization of adrenal nodules: can virtual unenhanced images replace true unenhanced acquisitions?.
        Abdom Radiol. 2021; 46: 4345-4352https://doi.org/10.1007/s00261-021-03062-3
        • Lennartz S
        • Pisuchpen N
        • Parakh A
        • et al.
        Virtual unenhanced images: qualitative and quantitative comparison between different dual-energy ct scanners in a patient and phantom study.
        Invest Radiol. 2022; 57: 52-61https://doi.org/10.1097/RLI.0000000000000802
        • Pourvaziri A
        • Parakh A
        • Mojtahed A
        • et al.
        Diagnostic performance of dual-energy CT and subtraction CT for renal lesion detection and characterization.
        Eur Radiol. 2019; 29: 6559-6570https://doi.org/10.1007/s00330-019-06224-6
        • Shaqdan KW
        • Parakh A
        • Kambadakone AR
        • et al.
        Role of dual energy CT to improve diagnosis of non-traumatic abdominal vascular emergencies.
        Abdom Radiol. 2019; 44: 406-421https://doi.org/10.1007/s00261-018-1741-7
        • Parakh A
        • Lennartz S
        • An C
        • et al.
        Dual-energy CT images: pearls and pitfalls.
        RadioGraphics. 2021; 41: 98-119https://doi.org/10.1148/rg.2021200102
        • Adam SZ
        • Rabinowich A
        • Kessner R
        • et al.
        Spectral CT of the abdomen: where are we now?.
        Insights Imaging. 2021; 12: 138https://doi.org/10.1186/s13244-021-01082-7
        • Mileto A
        • Ananthakrishnan L
        • Morgan DE
        • et al.
        Clinical implementation of dual-energy CT for gastrointestinal imaging.
        Am J Roentgenol. 2021; 217: 651-663https://doi.org/10.2214/AJR.20.25093
        • Kaltenbach B
        • Wichmann JL
        • Pfeifer S
        • et al.
        Iodine quantification to distinguish hepatic neuroendocrine tumor metastasis from hepatocellular carcinoma at dual-source dual-energy liver CT.
        Euro J Radiol. 2018; 105: 20-24https://doi.org/10.1016/j.ejrad.2018.05.019
        • Muenzel D
        • Lo GC
        • Yu HS
        • et al.
        Material density iodine images in dual-energy CT: detection and characterization of hypervascular liver lesions compared to magnetic resonance imaging.
        Euro J Radiol. 2017; 95: 300-306https://doi.org/10.1016/j.ejrad.2017.08.035
        • Shuman WP
        • Green DE
        • Busey JM
        • et al.
        Dual-energy liver CT: effect of monochromatic imaging on lesion detection, conspicuity, and contrast-to-noise ratio of hypervascular lesions on late arterial phase.
        AJR Am J Roentgenol. 2014; 203: 601-606https://doi.org/10.2214/AJR.13.11337
        • Toia GV
        • Mileto A
        • Wang CL
        • et al.
        Quantitative dual-energy CT techniques in the abdomen.
        Abdom Radiol. 2021; https://doi.org/10.1007/s00261-021-03266-7
        • Eberhard M
        • Mergen V
        • Higashigaito K
        • et al.
        Coronary calcium scoring with first generation dual-source photon-counting CT—first evidence from phantom and in-vivo scans.
        Diagnostics. 2021; 11: 1708https://doi.org/10.3390/diagnostics11091708
        • Rajendran K
        • Petersilka M
        • Henning A
        • et al.
        Full field-of-view, high-resolution, photon-counting detector CT: technical assessment and initial patient experience.
        Phys Med Biol. 2021; 66205019https://doi.org/10.1088/1361-6560/ac155e
        • Rajendran K
        • Petersilka M
        • Henning A
        • et al.
        First clinical photon-counting detector CT system: technical evaluation.
        Radiology. 2021; 212579https://doi.org/10.1148/radiol.212579
        • Euler A
        • Higashigaito K
        • Mergen V
        • et al.
        High-pitch photon-counting detector computed tomography angiography of the aorta: intraindividual comparison to energy-integrating detector computed tomography at equal radiation dose.
        Invest Radiol. 2022; 57: 115-121https://doi.org/10.1097/RLI.0000000000000816
        • Sartoretti T
        • Landsmann A
        • Nakhostin D
        • et al.
        Quantum iterative reconstruction for abdominal photon-counting detector CT improves image quality.
        Radiology. 2022; https://doi.org/10.1148/radiol.211931
        • Mergen V
        • Ried E
        • Allmendinger T
        • et al.
        Epicardial adipose tissue attenuation and fat attenuation index: phantom study and in-vivo measurements with photon-counting CT.
        Am J Roentgenol AJR. 2021; 21: 26930https://doi.org/10.2214/AJR.21.26930
        • Willemink MJ
        • Persson M
        • Pourmorteza A
        • et al.
        Photon-counting CT: technical principles and clinical prospects.
        Radiology. 2018; 289: 293-312https://doi.org/10.1148/radiol.2018172656
        • Euler A
        • Solomon J
        • Mazurowski MA
        • et al.
        How accurate and precise are CT based measurements of iodine concentration? A comparison of the minimum detectable concentration difference among single source and dual source dual energy CT in a phantom study.
        Eur Radiol. 2019; 29: 2069-2078https://doi.org/10.1007/s00330-018-5736-0
        • Jacobsen MC
        • Schellingerhout D
        • Wood CA
        • et al.
        Intermanufacturer comparison of dual-energy CT iodine quantification and monochromatic attenuation: a phantom study.
        Radiology. 2018; 287: 224-234https://doi.org/10.1148/radiol.2017170896
        • Pelgrim GJ
        • van Hamersvelt RW
        • Willemink MJ
        • et al.
        Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT.
        Eur Radiol. 2017; 27: 3904-3912https://doi.org/10.1007/s00330-017-4752-9
        • Schmidt C
        • Baessler B
        • Nakhostin D
        • et al.
        Dual-Energy CT-based iodine quantification in liver tumors – impact of scan-, patient-, and position-related factors.
        Acad Radiol. 2021; 28: 783-789https://doi.org/10.1016/j.acra.2020.04.021
        • Holz JA
        • Alkadhi H
        • Laukamp KR
        • et al.
        Quantitative accuracy of virtual non-contrast images derived from spectral detector computed tomography: an abdominal phantom study.
        Sci Rep. 2020; 10: 21575https://doi.org/10.1038/s41598-020-78518-5
        • Martin SS
        • Weidinger S
        • Czwikla R
        • et al.
        Iodine and fat quantification for differentiation of adrenal gland adenomas from metastases using third-generation dual-source dual-energy computed tomography.
        Invest Radiol. 2018; 53: 173-178https://doi.org/10.1097/RLI.0000000000000425
        • Mileto A
        • Marin D
        • Alfaro-Cordoba M
        • et al.
        Iodine quantification to distinguish clear cell from papillary renal cell carcinoma at dual-energy multidetector CT: a multireader diagnostic performance study.
        Radiology. 2014; 273: 813-820https://doi.org/10.1148/radiol.14140171
        • Cai LM
        • Hippe DS
        • Zamora DA
        • et al.
        A method for reducing variability across dual-energy CT manufacturers in quantification of low iodine content levels.
        Am J Roentgenol AJR. 2021; 21: 26714https://doi.org/10.2214/AJR.21.26714
        • Jiang X
        • Yang X
        • Hintenlang DE
        • et al.
        Effects of patient size and radiation dose on iodine quantification in dual-source dual-energy CT.
        Acad Radiol. 2021; 28: 96-105https://doi.org/10.1016/j.acra.2019.12.027
        • Große Hokamp N
        • Abdullayev N
        • Persigehl T
        • et al.
        Precision and reliability of liver iodine quantification from spectral detector CT: evidence from phantom and patient data.
        Eur Radiol. 2019; 29: 2098-2106https://doi.org/10.1007/s00330-018-5744-0
        • Mergen V
        • Sartoretti T
        • Klotz E
        • et al.
        Extracellular volume quantification with cardiac late enhancement scanning using dual-source photon-counting detector CT.
        Invest Radiol. 2022; https://doi.org/10.1097/RLI.0000000000000851